Repeatability of in-vitro optical quality measurements of intraocular lenses with a deflectometry technique effect of the toricity
Fecha de publicación:
Autores con afiliación ISPA
Otros autores
- Ferrer-Blasco, T
- Dominguez-Vicent, A
- Garcia-Lazaro, S
- Diez-Ajenjo, MA
- Esteve-Taboada, JJ
Grupos y Plataformas
Abstract
AIM: To evaluate the repeatability of an optical device for measuring the Zernike coefficients of toric intraocular lenses (IOLs) and assess whether its toricity has any impact in its repeatability. METHODS: An experienced technician used the NIMO TR1504 to measure the Zernike coefficients 30 times for an aperture of 4.50 mm for all lenses included. The IOLs included were divided into two group: toric and non-toric ones. The cylindrical powers of the toric lenses included in the present study were 1.00, 1.50, 2.25, 3.00 and 3.75 D. Finally, the repeatability of the NIMO TR1504 was described in terms of within subject standard deviation(Sw) and repeatability limit. RESULTS: The Sw was smaller than 0.011 mu m lens groups and all Zernike coefficients, and the difference between both groups was smaller than 0.004 mu m for all Zernike coefficients. Regarding the repeatability limit, this value was smaller than 0.025 mu m for the toric lens group and smaller than 0.031 mu m for the non toric lens one for all Zernike coefficients. Furthermore, the maximum difference both lenses groups was 0.010 mu m. CONCLUSION: The repeatability of the NIMO TR1504 to measure the optical quality is high and independent the lens toricity. These results reflect that this system is robust and could be used to measure the in-vitro optical quality of either toric or non toric IOLs.
Datos de la publicación
- ISSN/ISSNe:
- 2222-3959, 2227-4898
- Tipo:
- Article
- Páginas:
- 1139-1144
- PubMed:
- 30046530
International Journal of Ophthalmology IJO PRESS
Citas Recibidas en Web of Science: 1
Documentos
- No hay documentos
Filiaciones
Keywords
- in vitro measurements; optical quality; intraocular lens; repeatability
Cita
Ferrer Blasco, T; Dominguez Vicent, A; Garcia Lazaro, S; Diez Ajenjo, MA; Alfonso, JF; Esteve Taboada, JJ. Repeatability of in-vitro optical quality measurements of intraocular lenses with a deflectometry technique effect of the toricity. Int. J. Ophthalmol. 2018. 11. (7). p. 1139-1144. DOI: 10.18240/ijo.2018.07.11.