Posterior chamber collagen copolymer phakic intraocular lens with a central hole to correct myopia: One-year follow-up
Autores con afiliación ISPA
Otros autores
- Lisa, C
- Naveiras, M
- Alfonso-Bartolozzi, B
- Belda-Salmeron, L
- Montes-Mico, R
Abstract
PURPOSE: To assess the predictability, stability, efficacy, and safety of a newly developed posterior chamber phakic intraocular lens (pIOL) (model V4c Visian Implantable Collamer Lens) to correct moderate to high myopia. SETTING: Fernandez-Vega Ophthalmological Institute, Oviedo, Spain. DESIGN: Prospective case series. METHODS: This study evaluated eyes that had implantation of the new pIOL model with a central hole for myopia correction. Uncorrected (UDVA) and corrected (CDVA) distance visual acuities, refraction, intraocular pressure (IOP), endothelial cell density, pIOL vault, and adverse events were evaluated over 12 months. RESULTS: The study comprised 147 eyes of 80 patients. Preoperatively, the mean spherical equivalent (SE) was -8.80 diopters (D) +/- 2.60 (SD). At 12 months, the mean SE was -0.14 +/- 0.26 D, with 93.9% of eyes within +/- 0.50 D of the target and 100% of eyes within +/- 1.00 D. The mean UDVA and CDVA were 0.028 +/- 0.055 logMAR and 0.003 +/- 0.013 logMAR, respectively. The efficacy and safety indices were 1.00 and 1.04, respectively. All eyes had unchanged CDVA or gained 1 or more lines during the follow-up. The IOP remained stable over time; no eye developed anterior subcapsular cataract. Twelve months postoperatively, the mean vault was 405.5 +/- 184.7 mu m (range 100 to 980 mu m), and the mean endothelial cell loss was 1.7%. CONCLUSION: The good refractive and visual acuity outcomes and the highly stable IOP values obtained over 12 months support the use of the new pIOL model with a central hole for the correction of moderate to high myopia. (C) 2015 ASCRS and ESCRS
Datos de la publicación
- ISSN/ISSNe:
- 0886-3350, 1873-4502
- Tipo:
- Article
- Páginas:
- 1153-1159
- PubMed:
- 26189376
JOURNAL OF CATARACT AND REFRACTIVE SURGERY LIPPINCOTT WILLIAMS & WILKINS
Citas Recibidas en Web of Science: 43
Documentos
- No hay documentos
Filiaciones
Cita
Lisa, C; Naveiras, M; Alfonso Bartolozzi, B; Belda Salmeron, L; Montes Mico, R; Alfonso, JF. Posterior chamber collagen copolymer phakic intraocular lens with a central hole to correct myopia: One-year follow-up. J. Cataract. Refract. Surg. 2015. 41. (6). p. 1153-1159. DOI: 10.1016/j.jcrs.2014.10.030.
Portal de Investigación