Computer-assisted surgery improves rotational positioning of the femoral component but not the tibial component in total knee arthroplasty

Fecha de publicación:

Autores con afiliación ISPA

Otros autores

  • Noriega-Fernandez, A
  • Fernandez-Carreira, JM
  • Fernandez-Simon, JM
  • de los Rios, JL

Abstract

Computer-assisted surgery (CAS) may facilitate better positioning of total knee arthroplasty (TKA) along the coronal and lateral axes; however, there are doubts as to its usefulness in the rotational plane. This is a prospective study of 95 TKAs comparing two groups: the CAS group and the standard equipment group. The series comprises 95 cases. A radiography of the lower limb and computer tomographies (CTs) of the femoral condylar region, the proximal end of the tibia and the ankle were performed to measure rotational angulation. A month after TKA surgery, the radiography and the CTs were repeated to analyze the position of the prosthetic components in the rotational plane. In the coronal axis, both CAS and mechanical technique improved femoro-tibial alignment, but when there are preexisting deformities a parts per thousand yen4A degrees, CAS obtains better results. A strong correlation (R = 0.94, p = 0.001) was observed between the mean rotational axis measured with CT in the tibial plateau and that measured from the axis of the ankle. The mean initial femoral rotation of the complete series was 6.7A degrees and 2.7A degrees at 1-month follow-up (p < 0.001). In the standard instrumentation group, the femoral rotation went from 6.8A degrees to 2.3A degrees, whereas in the CAS group the femoral rotation went from 6.5A degrees to 3.1A degrees (p = 0.039), which is very close to the ideal 3A degrees angle of external rotation. Tibial rotation changed by 5.28A degrees for the entire patient population, but no differences were found when comparing CAS and standard instrumentation. CAS improves frontal alignment in TKA, especially in the presence of preoperative deformities. In the femoral component, navigation most closely replicated the ideal 3A degrees external rotation of the femoral component, but tibial rotation did not differ when comparing CAS to standard instrumentation. II.

Datos de la publicación

ISSN/ISSNe:
0942-2056, 1433-7347

KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY  WILEY

Tipo:
Article
Páginas:
3127-3134
PubMed:
25155048

Citas Recibidas en Web of Science: 28

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • Total knee arthroplasty; Navigation; Alignment; Femoral rotation; Tibial rotation

Cita

Compartir