Advanced nanocomposite materials for orthopaedic applications. I. A long-term in vitro wear study of zirconia-toughened alumina
Autores con afiliación ISPA
Otros autores
- Affatato, S
- Taddei, P
- Rocchi, M
- Fagnano, C
- Ciapetti, G
- Toni, A
Abstract
The use of ceramic-on-ceramic (alumina- and zirconia-based) couplings in hip joint prostheses has been reported to produce lower wear rates than other combinations (i.e., metal-on-polyethylene and ceramic-on-polyethylene). The addition of zirconia into an alumina matrix (zirconia-toughened alumina, ZTA) has been reported to result in an enhancement of flexural strength, fracture toughness, and fatigue resistance. The development of new processing routes in nonaqueous media has allowed to obtain high-density ZTA nanocomposites with a very homogeneous microstructure and a significantly smaller and narrower particle-size distribution of zirconia than conventional powder mixing methods. The aim of the present study was to set up and validate a new ZTA nanocomposite by testing its biocompatibillity and wear behavior in a hip-joint simulator in comparison with commercial alumina and experimental alumina specimens. The primary osteoblast proliferation onto ZTA nanocomposite samples was found to be not significantly different from that onto commercial alumina samples. After 7 million cycles, no significant differences were observed between the wear behaviors of the three sets of cups. In this light, it can be affirmed that ZTA nanocomposite materials can offer the option of improving the lifetime and reliability of ceramic joint prostheses. (c) 2005 Wiley Periodicals, Inc.
Datos de la publicación
- ISSN/ISSNe:
- 1552-4973, 1552-4981
- Tipo:
- Article
- Páginas:
- 76-82
- DOI:
- 10.1002/jbm.b.30462
- PubMed:
- 16292765
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS WILEY
Citas Recibidas en Web of Science: 41
Documentos
- No hay documentos
Filiaciones
Filiaciones no disponibles
Keywords
- zirconia-toughened alumina; ceramic materials; nanocomposites; hip simulator; biocompatibility
Cita
Affatato, S; Torrecillas, R; Taddei, P; Rocchi, M; Fagnano, C; Ciapetti, G; Toni, A. Advanced nanocomposite materials for orthopaedic applications. I. A long-term in vitro wear study of zirconia-toughened alumina. J. Biomed. Mater. Res. Part B. 2006. 78B. (1). p. 76-82. DOI: 10.1002/jbm.b.30462.