Determining the Alveolar Component of Nitric Oxide in Exhaled Air: Procedures and Reference Values for Healthy Persons

Fecha de publicación:

Autores con afiliación ISPA

  • Pere Casan Clarà

    Autor

Otros autores

  • Fortuna, AM
  • Balleza, M
  • Calaf, N
  • Gonzalez, M
  • Feixas, T

Abstract

Nitric oxide (NO) production has been described using a 2-compartment model for the synthesis and movement of NO in both the alveoli and the airways. The alveolar concentration of NO (CA(NO)), an indirect marker of the inflammatory state of the distal portions of the lung, can be deduced through exhalation at multiple flow rates. Our objective was to determine reference values for CA(NO). The fraction of exhaled NO (FENO) was measured in 33 healthy individuals at a rate of 50 mL/s: the subjects then exhaled at 10, 30, 100, and 200 mL/s to calculate CA(NO). A chemiluminescence analyzer (NIOX Aerocrine) was used to perform the measurements. The mean (SD) FENO was 15 (6) ppb. The mean CA(NO) was 3.04 (1.30) ppb. These values of CA(NO) measured in healthy individuals will allow us to analyze alveolar inflammatory behavior in respiratory and systemic processes. (C) 2008 SEPAR. Published by Elsevier Espana, S.L. All rights reserved.