Electrical impedance tomography: Standardizing the procedure in pneumology
Autores con afiliación ISPA
Otros autores
- de Lema, B
- Riu, P
Abstract
The following conditions are optimal for obtaining an adequate number of informative images by electric impedance tomography: a) patient seated or standing with hands at the nape of the neck; b) breathing at rest; c) recording of at least 300 images (at a frequency of 10 Hz), and d) readings taken at the sixth intercostal space.