CTF determination and correction in electron cryotomography

Fecha de publicación:

Autores con afiliación ISPA

Otros autores

  • Li, S
  • Crowther, RA

Abstract

Electron cryotomography (cryoET) has the potential to elucidate the structure of complex biological specimens at molecular resolution but technical and computational improvements are still needed. This work addresses the determination and correction of the contrast transfer function (CTF) of the electron microscope in cryoET. Our approach to CTF detection and defocus determination depends on strip-based periodogram averaging, extended throughout the tilt series to overcome the low contrast conditions found in cryoET. A method for CTF correction that deals with the defocus gradient in images of tilted specimens is also proposed. These approaches to CTF determination and correction have been applied here to several examples of cryoET of pleomorphic specimens and of single particles. CTF correction is essential for improving the resolution, particularly in those studies that combine cryoET with single particle averaging techniques. (c) 2006 Elsevier B.V. All rights reserved.

Datos de la publicación

ISSN/ISSNe:
0304-3991, 1879-2723

ULTRAMICROSCOPY  ELSEVIER

Tipo:
Article
Páginas:
587-596
PubMed:
16616422

Citas Recibidas en Web of Science: 134

Documentos

  • No hay documentos

Métricas

Filiaciones

Filiaciones no disponibles

Keywords

  • contrast transfer function; CTF; electron cryotomography; image processing algorithms; image reconstruction

Cita

Compartir