Agreement and Reproducibility of Anterior Chamber Angle Measurements between CASIA2 Built-In Software and Human Graders
Fecha de publicación:
Autores con afiliación ISPA
Otros autores
- Espinoza, Gustavo
- Iglesias, Katheriene
- Parra, Juan C.
- Serrano-Gomez, Sergio
- Prada, Angelica M.
- Galvis, Virgilio
Grupos y Plataformas
Abstract
Purpose: This study evaluated the agreement and reproducibility of ACA measurements obtained using the built-in software of the CASIA2 (Version 3G.1) and the measurements derived from expert clinicians. Methods: Healthy volunteers underwent ophthalmological evaluation and AS-OCT examination. ACA measurements derived from automated and manual SS location were obtained using the CASIA2 automated software and clinician identification, respectively. The intraobserver, interobserver reproducibility, CASIA2-human grader reproducibility and CASIA2 repeatability were assessed using intraclass correlation coefficients (ICCs). Results: The study examined 58 eyes of 30 participants. The CASIA2 software showed excellent repeatability for all ACA parameters (ICC > 0.84). Intraobserver, interobserver, and CASIA2-human grader reproducibility were also excellent (ICC > 0.87). Interobserver agreement was high, except for nasal TISA500, differing between observers 1 and 2 (p < 0.05). The agreement between CASIA2 measurements and human graders was high, except for nasal TISA500, where observer 1 values were smaller (p < 0.05). Conclusion: The CASIA2 built-in software reliably measures ACA parameters in healthy individuals, demonstrating high consistency. Although a small difference was observed in nasal TISA500 measurements, interobserver and CASIA2-human grader reproducibility remained excellent. Automated SS detection has the potential to facilitate evaluation and monitoring of primary angle closure disease.
Datos de la publicación
- ISSN/ISSNe:
- 2077-0383, 2077-0383
- Tipo:
- Article
- Páginas:
- -
- DOI:
- 10.3390/jcm12196381
- PubMed:
- 37835024
Documentos
- No hay documentos
Filiaciones
Keywords
- agreement; anterior chamber angle; AS-OCT; optical coherence tomography; primary angle closure glaucoma; reproducibility
Cita
Espinoza, G; Iglesias, K; Parra, JC; Rodriguez Una, I; Serrano Gomez, S; Prada, AM; Galvis, V. Agreement and Reproducibility of Anterior Chamber Angle Measurements between CASIA2 Built-In Software and Human Graders. J Clin Med. 2023. 12. (19). 6381. DOI: 10.3390/jcm12196381.
Portal de Investigación