Assessment of coagulation assays on Roche Cobas t711 analyzer: performance and clinical implications

Fecha de publicación:

Autores con afiliación ISPA

Otros autores

  • Ordóñez-Robles, M
  • Pons-Belda, OD
  • Moína, MJ

Grupos y Plataformas

Abstract

ObjectivesWe performed an analytical assessment of five coagulation tests [i.e. prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen, thrombin time (TT) and D-dimer] on the Roche Cobas t711 analyzer and a comparison study with the methodology in use at our laboratory (i.e. Werfen ACL Top 750 analyzer), expanding the analysis to the clinical implications of Cobas t711 implementation.MethodsImprecision studies were performed following the Clinical and Laboratory Standards Institute (CLSI) H57 A:2008 guideline. Linearity of D-dimer and fibrinogen tests was analysed according to the CLSI EP06-A: 2003 recommendations. For method comparison, the results were analyzed using the Bland-Altman plot and Passing-Bablok regression.ResultsImprecision met manufacturer claims for PT, aPTT and TT. D-dimer and fibrinogen tests showed a coefficient of variation (CV)% over manufacturer claims at certain concentration levels. Linearity ranges could not be verified. Comparison study revealed that results are not interchangeable for any test, a lower correlation for aPTT test and lower D-dimer results from Roche Cobas t711.ConclusionThe strength of this study relies on the analysis of the clinical implications of reporting Cobas t711 results compared to those obtained with the methodology in use at our laboratory. Different sensibility to factor deficiency, anticoagulant therapy and interferences might explain lower correlation rates obtained for the aPTT test. Different monoclonal antibodies used for D-dimer determination might explain the lower results obtained with the Cobas t711 analyzer. This aspect needs further studies given the relevance of D-dimer test to exclude thrombotic events and reinforces the need of harmonization in the haemostasis laboratory.

Datos de la publicación

ISSN/ISSNe:
0957-5235, 1473-5733

BLOOD COAGULATION & FIBRINOLYSIS  LIPPINCOTT WILLIAMS & WILKINS

Tipo:
Article
Páginas:
523-529
PubMed:
37823437

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • analytical performance; coagulation tests; Cobas t711; D-dimer; fibrinogen

Cita

Compartir