Early Hemodynamic Performance of the SAPIEN 3 Ultra Resilia and Evolut Valves: A Propensity-Matched Analysis

Fecha de publicación: Fecha Ahead of Print:

Autores con afiliación ISPA

Otros autores

  • Avvedimento, M
  • Mas-Peiro, S
  • Aktaa, S
  • Nombela-Franco, L
  • Gonzalez, IC
  • Franzone, A
  • Campelo-Parada, F
  • Urena, M
  • Cepas-Guillén, P
  • Tirado, G
  • Laffond, AE
  • Esposito, G
  • Tamir, M
  • Suc, G
  • Mohammadi, S
  • Pibarot, P
  • Côté, M
  • Webb, JG
  • Rodés-Cabau, J

Grupos y Plataformas

Abstract

Background: Limited data are available on early hemodynamic outcomes of the balloon-expandable SAPIEN 3 Ultra Resilia (S3UR) (Edwards Lifesciences, Irvine, CA) compared with the self-expandable Evolut (Medtronic, Minneapolis, MN) transcatheter heart valves. In this study we aimed to compare the hemodynamic performance of the S3UR with contemporary Evolut platforms (Evolut PRO, PRO+, FX) in a propensity-matched study. Methods: A total of 307 patients who received an S3UR valve were matched with 488 patients who received Evolut valves (mean age, 80.1 +/- 7 years; 46.2% women). The primary end point was valve hemodynamics (aortic mean gradient, effective orifice area [EOA], severe prosthesis-patient mismatch [PPM], and moderate to severe aortic regurgitation [AR]) at 1-3 months after transcatheter aortic valve replacement. Secondary end points included clinical outcomes. Results: Evolut valves were associated with lower mean gradients (8.0 +/- 3.1 mm Hg vs 9.8 +/- 3.8 mm Hg; P < 0.001) but higher rates of residual AR (mild AR, 10.7% vs 4.1%; moderate to severe AR, 4.2% vs 1.4%; P < 0.001). EOA values and PPM rates were comparable. In patients with small aortic annulus (< 430 mm(2)), Evolut valves showed larger EOA (1.94 +/- 0.3 cm(2) vs 1.83 +/- 0.4 cm(2); P = 0.038) and lower gradients (7.8 +/- 2.4 mmHg vs 10.9 +/- 4.5 mmHg; P < 0.010). Moderate to severe AR and PPM were similar between groups. Despite comparable device success, early safety outcomes favoured the S3UR (81.8% vs 60.5%; P < 0.010 for S3UR and Evolut, respectively), because of higher rates of pacemaker implantation, atrial fibrillation, major vascular complication, and type 2-3 bleeding in the Evolut cohort. Conclusions: Evolut valves showed lower transprosthetic gradients but a higher rate of residual AR compared with the S3UR. Further studies are needed to validate these findings and investigate their effect on long-term outcomes.

Copyright © 2025 Canadian Cardiovascular Society. Published by Elsevier Inc. All rights reserved.

Datos de la publicación

ISSN/ISSNe:
0828-282X, 1916-7075

CANADIAN JOURNAL OF CARDIOLOGY  ELSEVIER SCIENCE INC

Tipo:
Article
Páginas:
2407-2416
PubMed:
40812524

Citas Recibidas en Web of Science: 4

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • Aortic stenosis; transcatheter heart valve; transcatheter aortic valve replacement; SAPIEN 3 Ultra Resilia; Evolut