Contractility sensor-guided optimization of cardiac resynchronization therapy: results from the RESPOND-CRT trial

Fecha de publicación:

Autores con afiliación ISPA

Otros autores

  • Brugada, J
  • Delnoy, PP
  • Brachmann, J
  • Reynolds, D
  • Padeletti, L
  • Noelker, G
  • Kantipudi, C
  • Dichtl, W
  • Borri-Brunetto, A
  • Verhees, L
  • Ritter, P
  • Singh, JP
  • RESPOND CRT Investigators

Abstract

Aims Although cardiac resynchronization therapy (CRT) is effective in patients with systolic heart failure (HF) and a wide QRS interval, a substantial proportion of patients remain non-responsive. The SonR contractility sensor embedded in the right atrial lead enables individualized automatic optimization of the atrioventricular (AV) and interventricular (VV) timings. The RESPOND-CRT study investigated the safety and efficacy of the contractility sensor system in HF patients undergoing CRT. Methods and results RESPOND-CRT was a prospective, randomized, double-blinded, multicentre, non-inferiority trial. Patients were randomized (2: 1, respectively) to receive weekly, automatic CRT optimization with SonR vs. an Echo-guided optimization of AV and VV timings. The primary efficacy endpoint was the rate of clinical responders (patients alive, without adjudicated HF-related events, with improvement in New York Heart Association class or quality of life), at 12 months. The study randomized 998 patients. Responder rates were 75.0% in the SonR arm and 70.4% in the Echo arm (mean difference, 4.6%; 95% CI, -1.4% to 10.6%; P<0.001 for non-inferiority margin -10.0%) (Table 2). At an overall mean follow-up of 5486190 days SonR was associated with a 35% risk reduction in HF hospitalization (hazard ratio, 0.65; 95% CI, 0.46-0.92; log-rank P = 0.01). Conclusion Automatic AV and VV optimization using the contractility sensor was safe and as effective as Echo-guided AV and VV optimization in increasing response to CRT.

Datos de la publicación

ISSN/ISSNe:
0195-668X, 1522-9645

EUROPEAN HEART JOURNAL  OXFORD UNIV PRESS

Tipo:
Article
Páginas:
730-738
PubMed:
27941020

Citas Recibidas en Web of Science: 75

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • SonR; AV and VV optimization; CRT optimization; Cardiac contractility; RESPOND; HF hospitalization; Adaptive CRT

Cita

Compartir