A high-performance liquid chromatographic assay with diode-array detection has been developed for the in-process control of zolpidem synthesis and for the analysis of the drug and its synthetic intermediates. The separation uses a 4.6 mm i.d. reversed-phase Kromasil C-18 (150 mm) column, 5 mum particle size with a gradient elution mode of acetonitrile and 0.02 M NH4OAc (adjusted to pH 8.0) as the mobile phase (flow rate 1.0 ml min(-1)). The analysis is performed in 12 min. The method is simple, rapid and highly specific. (C) 2004 Elsevier B.V. All rights reserved.
Datos de la publicación
- ISSN/ISSNe:
- 0731-7085, 1873-264X
- Tipo:
- Article
- Páginas:
- 925-928
- PubMed:
- 15533692
Journal of Pharmaceutical and Biomedical Analysis ELSEVIER
Citas Recibidas en Web of Science: 11
Documentos
- No hay documentos
Filiaciones
Filiaciones no disponibles
Keywords
- zolpidem; in-process control; HPLC
Cita
Laviana, L; Mangas, C; Fernandez Mari, F; Bayod, M; Blanco, D. Determination and in-process control of zolpidem synthesis by high-performance liquid chromatography. J Pharm Biomed Anal. 2004. 36. (4). p. 925-928. DOI: 10.1016/j.jpba.2004.08.025.