Amperometric glutamate biosensor based on poly(o-phenylenediamine) film electrogenerated onto modified carbon paste electrodes
Autores con afiliación ISPA
Otros autores
- Alvarez-Crespo, SL
- Miranda-Ordieres, AJ
- Tunon-Blanco, P
Abstract
A glutamate biosensor was developed by electropolymerizing o-phenylenediamine on a dehydrogenase-NAD(+) modified carbon paste electrode. The amperometric response to glutamate was obtained by means of the electrocatalytic oxidation of the enzymatically produced NADH, at an applied potential close to OV (Ag/AgCl). After optimizing carbon paste composition, polymer thickness and operating variables, a linear amperometric response to glutamate was obtained within the concentration range 5.0 x 10(-6)-7.8 x 10(-5) M with a detection limit of 3.8 x 10(-6) M. The biosensor was applied to the determination of glutamate in chicken bouillon cubes. Good accuracy was found with respect to a reference enzymatic spectrophotometric method. (C) 1997 Elsevier Science Limited.
Datos de la publicación
- ISSN/ISSNe:
- 0956-5663, 1873-4235
- Tipo:
- Article
- Páginas:
- 739-747
- PubMed:
- 9421885
BIOSENSORS & BIOELECTRONICS ELSEVIER ADVANCED TECHNOLOGY
Citas Recibidas en Web of Science: 51
Documentos
- No hay documentos
Filiaciones
Filiaciones no disponibles
Keywords
- glutamate; poly(O-phenylenediamine) amperometric biosensor; carbon paste
Cita
Alvarez Crespo, SL; Lobo Castanon, MJ; Miranda Ordieres, AJ; Tunon Blanco, P. Amperometric glutamate biosensor based on poly(o-phenylenediamine) film electrogenerated onto modified carbon paste electrodes. Biosens. Bioelectron. 1997. 12. (8). p. 739-747. DOI: 10.1016/S0956-5663(97)00041-9.
Portal de Investigación