Amperometric glutamate biosensor based on poly(o-phenylenediamine) film electrogenerated onto modified carbon paste electrodes

Fecha de publicación:

Autores con afiliación ISPA

Otros autores

  • Alvarez-Crespo, SL
  • Miranda-Ordieres, AJ
  • Tunon-Blanco, P

Abstract

A glutamate biosensor was developed by electropolymerizing o-phenylenediamine on a dehydrogenase-NAD(+) modified carbon paste electrode. The amperometric response to glutamate was obtained by means of the electrocatalytic oxidation of the enzymatically produced NADH, at an applied potential close to OV (Ag/AgCl). After optimizing carbon paste composition, polymer thickness and operating variables, a linear amperometric response to glutamate was obtained within the concentration range 5.0 x 10(-6)-7.8 x 10(-5) M with a detection limit of 3.8 x 10(-6) M. The biosensor was applied to the determination of glutamate in chicken bouillon cubes. Good accuracy was found with respect to a reference enzymatic spectrophotometric method. (C) 1997 Elsevier Science Limited.