Evaluation of uncertainty sources in the determination of testosterone in urine by calibration-based and isotope dilution quantification using ultra high performance liquid chromatography tandem mass spectrometry

Fecha de publicación:

Autores con afiliación ISPA

Otros autores

  • Pitarch-Motellon, J
  • Roig-Navarro, AF
  • Sancho, JV
  • Ibanez, M
  • Fabregat-Cabello, N
  • Pozo, OJ
  • Ventura, R
  • Gago, AG
  • Artabe, AE
  • Van Eenoo, P
  • Deventer, K
  • Dehnes, Y
  • Rzeppa, S

Grupos y Plataformas

Abstract

Three quantification methodologies, namely calibration with internal standard (Cal-IS, non-weighted), weighted calibration with internal standard (wCal-IS) and isotope pattern deconvolution (IPD) have been used for the determination of testosterone in urine by LC-MS/MS. Uncertainty has been calculated and compared for the three methodologies through intra- and inter-laboratory reproducibility assays. IPD showed the best performance for the intra-laboratory reproducibility, with RSD and combined uncertainty values below 4% and 9% respectively. wCal-IS showed similar performance, while Cal-IS where not constant and clearly worse at the lowest concentration assayed (2 ng/mL) reaching RSD values up to 16%. The inter-laboratory assay indicated similar results although wCal-IS RSD (20%) was higher than IPD (10%) and Cal-IS get worse with RSD higher than 40% for the lowest concentration level. Uncertainty budgets calculated for the three procedures revealed that intercept and slope were the most important factors contributing to uncertainty for Cal-IS. The main factors for wCal-IS and IPD were the volumes of sample and/or standard measured. (C) 2017 Elsevier B.V. All rights reserved.

Datos de la publicación

ISSN/ISSNe:
0021-9673, 1873-3778

JOURNAL OF CHROMATOGRAPHY A  ELSEVIER

Tipo:
Article
Páginas:
73-80
PubMed:
28601363

Citas Recibidas en Web of Science: 10

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • Testosterone; urine; LC-MS/MS; uncertainty; IDMS; Isotope pattern deconvolution