Determination of Polychlorinated Biphenyls in Solid Samples by Isotope Dilution Mass Spectrometry Using Cl-37-Labeled Analogues

Fecha de publicación:

Autores con afiliación ISPA

Otros autores

  • Somoano-Blanco, L
  • Fonseca, SG

Abstract

This work describes the first application of Cl-37-labeled compounds to isotope dilution mass spectrometry (IDMS). The synthesis of 12 Cl-37-labeled polychlorinated biphenyls (PCBs) was carried out by the chlorination of biphenyl with isotopically enriched chlorine gas, generated by the direct oxidation of (NaCl)-Cl-37 with potassium peroxymonosulfate. After an exhaustive purification due to the presence of other congeners, the concentration and the isotopic enrichment of all Cl-37-labeled PCBs in the mixture was determined. The proposed procedure allows the simultaneous quantification of every isotope diluted PCB congener in a single gas chromatography-tandem mass spectrometry (GC-MS/MS) injection without resorting to a methodological calibration graph. The results obtained here demonstrate that the use of Cl-37-labeled analogues provides results in agreement with the certified values of three different Certified Reference Materials (marine sediment SRM 1944, fish tissue 1947, and loamy soil CRM 962-50) and analytical figures of merit comparable to those obtained using regular IDMS procedures based on the use of commercially available C-13-labeled analogues.